Is medical device packaging exempt from PPWR?

Verified on 20 November 2026 against Regulation (EU) 2025/40 · Relies on: Articles 3(1)(49), 5, 6(11), 6(12), 7(4), 10, 12(11), 15, 24, 44, 45

No, not fully. Contact-sensitive packaging of medical devices and in vitro diagnostic devices is taken out of two rules only: recyclability (Article 6(11)(b) and (c)) and minimum recycled plastic content (Article 7(4)(b) and (c)). Every other PPWR duty applies, including the technical file and signed Declaration of Conformity, the heavy metals limit, packaging minimisation, empty-space rules and producer registration and fees. Packaging that does not count as contact-sensitive is not relieved from anything.

What is lifted and what is not

RequirementContact-sensitive device packagingWhen it appliesArticle
RecyclabilityDoes not apply—Art. 6(11)(b), (c)
Minimum recycled plastic contentDoes not apply (also covers research-use and investigational devices)—Art. 7(4)(b), (c)
Technical documentation and Declaration of ConformityAppliesNowArt. 15, 38, 39
Heavy metals, 100 mg/kgAppliesNowArt. 5(4)
Producer registration and EPR feesAppliesNowArt. 44, 45
Empty space in sales packaging kept to the minimumApplies12 Feb 2028Art. 24(4)
Packaging minimisationApplies1 Jan 2030Art. 10
Harmonised waste-sorting labelCommission guidance: not for devices used only by professionals. Space-based relief for immediate and outer packaging2028 or laterArt. 12(1), 12(11)

Two more device-specific points:

Why so many teams believe they are exempt

Early summaries said "medical devices are exempt." That was shorthand for the two exemptions above. The regulation's scope covers all packaging, whatever the product (Art. 1, 3(1)(1)). Nothing removes device packaging from scope.

The exemptions still create paperwork

The Commission's FAQ says exemptions must be assessed and recorded in the technical documentation. For each pack you claim relief on, write down:

The exemptions are not permanent by design

ReviewDateArticle
Commission checks whether new recycled-content derogations or changes to the exemption list are neededBy 1 January 2028Art. 7(12)
Commission reviews the recycled content rules and the relevance of the exemptionsBy 12 February 2032Art. 7(14)
Commission reviews the recyclability exemptions, including for devicesBy 1 January 2035Art. 6(12)

A pack designed today will often still be on the market at the 2035 review. Plan for the chance that the relief narrows.

What to do now

  1. List every pack level for each device: sterile barrier, protective packaging, sales carton, shipper, pallet.
  2. Mark which are contact-sensitive, with your reason.
  3. Build the technical file and Declaration for every pack, exempt or not.
  4. Treat non-exempt outer and transport packs as fully in scope for 2028 and 2030.

Want every pack level classified and documented?

Book a medical device Exposure Scan

Sources: Regulation (EU) 2025/40 (OJ L, 22.1.2025); European Commission guidance document, 30 March 2026; DG ENV PPWR FAQ, March 2026. General information, not legal advice.